Autoclave and sterilizer repair. We pick it up and leave you a loaner.
Midmark, Tuttnauer and SciCan Statim tabletop autoclaves, repaired on our bench and tested before they come back. Our courier takes your unit and swaps in a tested loaner on the same stop, so the practice never stops sterilizing.
The tabletop units practices actually run
Most failures are the same handful of parts: a hardened door gasket, a heating element, a sticking valve, a dirty water sensor, a tired board. We find which one on the bench, quote it and fix it.
Midmark and Ritter
M11, M9, M3 and older UltraClaves: gaskets, heaters, valves, boards, C-codes.
Tuttnauer
2340M and 2540M, EZ9, EZ10 and EZ11, 3870, Elara 11, T-Edge: LOW WATER, LOW HEAT and the rest.
SciCan Statim
Statim 2000 and 5000, G4 included: cassette seals, pumps, steam generators, cycle faults.
SciCan Bravo
Bravo G4 Class B vacuum units: vacuum faults, door gaskets, filters.
Tabletop steam sterilizers only — the countertop units dental, medical, veterinary, tattoo and nail practices use. Floor-standing and hospital sterilizers are a different job and we do not take them.
Swapped in on the same stop, spore-tested before it leaves
Most repair services ship you a loaner and charge you to send it back by a deadline. Our courier brings it with him: he takes your unit and leaves the loaner, then swaps them back when yours is repaired.
- $149 for an M9-class unit, $199 for an M11-class or Statim, up to 21 days
- Free on repairs over $900, on overhauls and on maintenance plans
- Every loaner leaves with a passing spore-test record and a cycle printout
Four steps
Tell us the fault
Send the form or text us the model and the error on the screen. A technician replies with a price range.
Pickup and loaner
The courier collects the unit on the next run in your area and leaves a loaner if you need one. $125 covers pickup, diagnostic and return, credited toward the repair.
Diagnose and quote
We find the fault on the bench and text you the price. Nothing is done until you approve it.
Tested and returned
Full cycles on the bench with temperature and pressure checked, the printout for your log, and the unit back on your counter.
Back in service the way California expects
Dental offices in California spore-test at least weekly (16 CCR §1005). Body-art studios must spore-test after any major repair and at least monthly. We test every unit on full cycles before it leaves and send the printout — then you run your biological indicator on the first load back and log it.
What we test after a repair → · Spore testing in California →
What the code on your screen means
Some faults you can clear yourself: a unit that is not level, the wrong water, a dirty sensor. Check here before you book.
Midmark error codes
How to read a Midmark C-code, plus E-codes and the new-style codes.
Midmark C983
What C983 means on an M9 or M11 and what fixes it.
Tuttnauer error messages
LOW WATER, LOW HEAT, HI TEMP and the rest, explained.
Statim cycle faults
Every Statim cycle fault number and what it points to.
Statim cycle fault 3
The most common Statim fault and how to clear it.
Weekly courier runs across Southern California
- Los Angeles County
- Orange County
- San Diego County
- Inland Empire
- Ventura County
- Santa Barbara County
Same unit, different rules
Medical clinics
Urgent care, dermatology, OB/GYN, ENT and eye clinics.
Podiatry
High cycle counts, wet cassettes and rust on nippers.
Dental offices
Weekly spore tests, a loaner, and repairs that keep the chair schedule.
Tattoo and piercing studios
Tattoo, piercing and permanent makeup: body-art rules and one sterilizer.
Veterinary clinics
Surgical packs, long cycles and chamber autoclaves.
Questions
Do you repair autoclaves on site?
No — we pick the unit up and repair it on our bench in West Hollywood, where the parts, the test equipment and the time to do it properly are. A loaner can be swapped in on the same stop, so you keep sterilizing while yours is out.
How much does an autoclave repair cost?
Pickup, bench diagnostic and return is $125, credited toward the repair. A door gasket is from $249, a heating element from $449, a full overhaul $1,295–$1,595. You get a written quote before any work.
How long will we be without our sterilizer?
Typical bench time is 3 to 5 business days on the bench, parts permitting. With a loaner you are not without one at all: the courier swaps it in when he takes yours, and swaps back when he returns it.
Which brands do you repair?
Midmark and Ritter, Tuttnauer, SciCan Statim and Bravo, and older Pelton & Crane units — the tabletop sterilizers most dental, medical, veterinary and tattoo practices run.
What happens after the repair?
Every unit runs full cycles on the bench with temperature and pressure checked against its settings before it leaves. You get the printout or log for your records, and we remind you to run your own spore test on the first load back, as California practices are expected to.
If you decline the quote, the $125 covers the pickup, the bench diagnostic and bringing the unit back, and that is all you pay.
Book a pickup
Tell us the model, the error on the screen and what it does. A technician reads it and texts you back with a price range and the next pickup day in your area.
- Pickup, bench diagnostic and return — $125, credited toward the repair
- A tested loaner swapped in on the same stop, from $149 — free on repairs over $900
- A written quote before any work starts
- Typical bench time: 3 to 5 business days on the bench, parts permitting
Rather talk? Call or text (424) 622-8416.